World Aquaculture Singapore 2026

June 2-5, 2026

Singapore

APPLICATION OF AN AUTOMATED SEQUENTIAL INJECTION ANALYSIS SYSTEM FOR REAL-TIME AMMONIA MONITORING IN AQUACULTURE

Hung-Yu Chen , Tzong-Jih Cheng *

Department of Biochemical Science & Technology, College of Life Science, National Taiwan University, Taipei, Taiwan

D13631001@ntu.edu.tw , tzongjih@ntu.edu.tw



ABSTRACT

Ammonia nitrogen (NH3/NH4+) management is vital for aquaculture. This study presents an automated monitoring platform based on Sequential Injection Analysis (SIA) with fluorescence detection. The system utilizes a sandwich-like injection method and automated in-line pre-treatment for efficient in-situ reaction. Results show high repeatability (RSD<5% ) and reproducibility(RSD<8%), offering a robust tool for precision water quality management.

1. BACKGROUND AND METHODOLOGY

Conventional manual monitoring is labor-intensive. Our SIA system automates the analytical cycle using a precision manifold. A "sandwich technology-liked method" stacks reagents and samples for optimal mixing. The sequence includes an automated in-line pre-treatment and heating phase (0~420 s) to accelerate fluorogenic kinetics. An air-bubble-assisted cleaning cycle ensures baseline stability between runs.

2. RESULTS AND DISCUSSION

Overlaid chronograms show remarkable consistency. Isolated bubble artifacts (~50 s) are effectively distinguished from analyte signals. Quantitative results (Fig 2) show a significant response (0.59V vs. 0.27V blank), confirming high reproducibility.

3. CONCLUSION

The developed SIA system ensures data integrity via in-line heating and sandwich-like injection, providing a cost-effective, automated solution for sustainable fishery management.